Aoli Medical Instrument
Add:Tech-industry Developping Area, Jiangshan, Ningbo, China
Tel:0574-88098138
Fax:0574-88098266
http://www.nbaoli.com
E-mail:nbaoli@nbaoli.com |
|
 |
 |
| |
|
 |
| ECG-8903 Three Channel Electrocardiograph(3.5 |
|
Primary Technique Index
| Lead system |
12 leads, 3 channels sampling synchronized |
| Input mode |
Floating and Eliminates trembling protects |
| Input impedance |
>50M Ω |
| Output impedance |
< 100 Ω |
| Patient leak current |
< 10 µA |
| Input loop current |
≤ 0.1 µA |
| Calibration voltage |
1mV±1% |
| Output voltage |
0.5mm/mV±3% |
| A/D conversion |
14 bit |
| Frequency response |
0.05Hz-150Hz |
| Time constant |
>3.2s |
| Common-mode restrain ratio |
>80dB >100dB(with AC filter) |
| EMG filter |
25Hz(-3dB)/35Hz(-3dB)/35Hz(-3dB) |
| Standard sensitivity |
10mm/mV±0.2mm/mv |
| Sensitivity selection |
2.5、5、10、20mm/mV, Transformation error±3% |
| Polarization-endured voltage |
±400mV |
| Noise level |
≤ 15 µVp-p |
| Baseline control |
Auto-adjusting |
| ECG output |
0.5V/mV±5%, single port= 100 Ω |
| Power supply |
Ac: 200V, 50Hz, 15VA Dc: 12V rechargeable battery |
| safety standard |
IEC I category, CF style |
| Operation method |
Auto or Manual |
| Analysis function |
ECG parameters auto-measure, auto-analysis |
| Printing speed |
12.5、25、50mm/s±1% |
| Recording system |
Thermal-array printing system |
| Recording paper standard |
high-speed rool 50mmX30m |
| Dimensions |
(L x W x H)310X230X70mm |
| Packing size |
(L x W x H)400X360X130mm |
| Net weight |
2.7kg |
| Gross weight |
4.5kg |
|
| |
|
| [返回] |
|
|